Product code | SS 20.515 / 516 | SS 20.400 / SS 23.400 | SS 20.415 | SS 20.500 |
Application | Filter monitoring, Room monitoring | Flow direction (turbulence degree) | Flow direction (turbulence degree) | Filter monitoring, Room monitoring if corrosive medias are present |
Sensor head | Dumb-bell head | Thermopile head | Thermopile head | Dumb-bell head |
Measuring range | 0,06 … 2,5 m/s | 0,2 … 20 m/s | 0,05 … 20 m/s | 0,06 … 35 m/s |
Medium temperature | -20 ... +85 °C | 0 … +60 °C | 0 … +60 °C | -20 … +85 °C |
Moistness area | 0 ... 95 % rF | 0 ... 95 % rF | 0 ... 95 % rF | 0 ... 95 % rF |
Specials |
SS 20.515 for disinfection with |
Bi-directional flow sensor Switching outputs Measurement of turbulence degree Self monitoring function Detection of contamination ATEX version for Ex-secured areas |
Disinfection with H2O2 and alcohols Bi-directional flow sensor Switching outputs Measurement of turbulence degree Self monitoring function Detection of contamination |
Disinfection with H2O2 and alcohols Protection against corrosion e.g.: - hydrochloric acid - caustic soda - acetone High precision adjustment |
Mounting | Sealing (stainless steel) - Counternut - Screw adapter - Shanknut - Flange bushing |
Easy one-hole-mounting | Screwed bush M25 (stainless steel) Quick mounting system - Counternut - Screw adapter - Shank nut - Flange bushing |
Mounting lengths from 100 to 1.000 mm available |
Output signals | 0 ... 10 V 0 ... 20 mA 4 ... 20 mA |
0 ... 2 V 0 ... 5 V 0 ... 10 V 0 ... 20 mA 4 ... 20 mA |
0 ... 2 V 0 ... 5 V 0 ... 10 V 0 ... 20 mA 4 ... 20 mA |
Auto UI 4 … 20 mA ↔ 0 … 10 V |
Application examples:
Cleanroom supervision – monitoring
Energy saving and efficiency through
Higher measurement accuracy for efficient reduction of the flow velocity beyond production times (higher accuracy compared to stationed pressure measurement)
Plant monitoring
Pharmaceutical Industry
Semi-conductor Industry